Flotation catheter or Swan Ganz . It is one of the most used tools in the critical patient.
Summary
[ hide ]
- 1 Features
- 1 Internal
- 2 Indications
- 3 Typologies
- 4 Insertion technique
- 1 Problems related to insertion
- 5 Complications
- 6 See also
- 7 Sources
characteristics
It is inserted into a thick caliber vein, either by percutaneous puncture through the internal, subclavian, or femoral jugular, or by dissection in the antecubital area. It is used to measure pulmonary pressure and wedge pressure of the pulmonary artery that reflects the pressure of the left atrium and the final diastolic pressure of the left ventricle . The catheter is inserted into the right atrium, the balloon is inflated, and the catheter follows blood flow through the tricuspid valve into the right ventricle and the pulmonary artery.
It allows the measurements of the pressure in the pulmonary vascular tree and the sampling of mixed venous blood:
- Right atrium .
- Pulmonary artery
- Pulmonary (wedge) capillary or pulmonary artery occlusion pressure (PCP).
- Determination of cardiac output [[(GC) by thermodilution .
- Obtaining mixed venous blood samples.
Internal
Inside there are four separate lights:
- Thermistor .
- Via Do Balao .
- Distal route (yellow).
- Proximal route ( Blue ).
Indications
- For diagnostic purposes:
- Pulmonary edema .
- Types of shock .
- Pulmonary hypertension.
- Acute mitral regurgitation.
- Ventricular septum defects (detection of an oximetric jump between AD and AP ).
- Infarction of the right ventricle .
- Cardiac tamponade.
- Arrhythmias .
- Carcinomatous lymphangitis ( microvascular cytology of blood drawn from the distal orifice with the balloon swollen and nailed).
- Microvascular thrombosis ( pulmonary angiography with contrast injection through the distal orifice and the swollen and nailed balloon).
- Monitoring and evaluation of:
Treatments in critically ill patients with severe lung and hemodynamic diseases.
Typologies
- Pediatric (thinner)
- With additional light to infuse liquids.
- With fiber optic bundles for SvO2 monitoring .
- With special temperature sensor , fast response to measure RV ejection fraction.
- With system for continuous measurement of cardiac output.
- With additional light with outlet at RV level to measure its pressure or to introduce a pacemaker electrocatheter .
- Others to record electrical activity in AD and VD.
Insertion technique
- Continue advancing until obstructing the AP branch (nailing), observing an atrial-type curve, disappearance of differential pressure and decrease in mean pressure.
- Deflate the balloon with reappearance of the AP curve (never remove the catheter with an inflated balloon).
- Suture the introducer to the skin to fix the position and extend and fix the protective sleeve for possible manipulations.
- Radiological verification.
Insertion issues
Non-visualization of the RV curve after inserting the catheter 30-40 [[cm], responds to an AD coil or passage to the inferior or internal jugular cava .
-
- Deflate the balloon, remove it up to 20 cm and repeat the operation.
If after viewing the RV curve, more than 15 cm have been inserted and the AP curve is not obtained, it is due to catheter coiling.
-
- Deflate the balloon, withdraw it to AD and repeat the operation (rotate the catheter clockwise, raise the head of the bed or tilt the patient).
-
- If it fails, carry out introduction with radio control.
If a PCP-like image is obtained immediately after the RV, it is because the tip of the catheter has been placed at the level of the coronary sinus or below the pulmonary valve.
-
- The AD curve is always followed by the VD curve, and this by the AP curve and after the PCP curve.
Difficult identification of pressure curves due to hypovolemia (decrease in the differences between the RV and AP diastolic pressures, and between the mean pressure of AP and PCP), due to an increase in RV end-diastolic pressure (the transition from RV to AP) can be difficult to appreciate).
Complications
- Pulmonary infarction:
- It occurs if the catheter migrates distally to the occlusion position and is not detected.
- Systematic need to continuously monitor PAP.
- Lung joint rupture:
- Excessive inflation of the ball.
- Inflated the ball when it is locked.
- Prolonged maintenance in the locked position.
- Advancement of the catheter with the balloon deflated or washed too vigorously.
- Low incidence: 0.06 – 0.7 % , but it is associated with high mortality (50%).