Analytical balance

Analytical balance. The analytical balance is the instrument most used by the chemist, since by means of it it is possible to know exactly: the mass of matrix destined for analysis, the mass of substances to prepare solutions of exact concentration, the mass of precipitates in gravimetric analysis .

Summary

[ hide ]

  • 1 Description
  • 2 Features
  • 3 Components
  • 4 Installation requirements
  • 5 Maintenance
  • 6 Sources

Description

The balance is an instrument that measures the mass of a body or substance, using the force of gravity acting on the body as a means of comparison. Depending on the work to be carried out, the most suitable type of balance is selected in terms of sensitivity and speed in weighing. The sensitivity of a balance depends on its capacity: a balance designed to weigh kilograms will hardly have the necessary sensitivity to have reproducibility in weighing milligrams. In the laboratory, the balance is used to carry out quality control activities –with devices such as pipettes-, to prepare mixtures of components in predefined proportions and to determine specific gravity or weights.

Modern analytical balances, which can offer reading precision values ​​from 0.1 micron to 0.1 mg, are well developed so that the use of special rooms for weight measurement is not necessary. Still, the simple use of electronic circuits does not eliminate the interactions of the system with the environment. Of these, the physical effects are the most important because they cannot be suppressed.

The scales differ from each other by the design, the principles used and the metrology criteria they use. At present, it could be considered that there are two main groups: mechanical scales and electronic scales.

One of the demands placed on quantitative analytical chemistry is Accuracy , that is, that the results be as close to reality as possible. Of course, this accuracy should generally be achieved with very small Sample masses , since working with large Matrix masses is cumbersome and time consuming, therefore the desired accuracy can only be achieved with an instrument capable of achieving high levels of accuracy in weighing of small masses. These objectives are achievable in practice with the use of the analytical balance.

characteristics

The analytical balance used in chemical analysis generally allows weighing masses less than 200 grams with a sensitivity of 0.1 mg and in some cases 0.01 mg, that is, it is capable of weighing substances, reporting values ​​up to the fourth or fifth decimal place. Another important characteristic of the analytical balance is its fidelity (Accuracy), consisting of the ability to give the same value every time the same object is weighed several consecutive times.

Components

One of the systems with which it operates is the so-called “Support System” . This is a method by which all weighing objects can be held. It is made up of a series of elements whose operation is important to know to operate the meter, these are three: the stage, a series of stirrups and the column, there is a second analytical balance management system called “Balance system” . It is a sensitive procedure that has the function of recording the weight of the object using a scale method. It is made up of a cross and some blades, the third system is the “Scale system”. You are tasked with dropping the weight value, but always after the balance system registers the weight of the object. It consists of an optical scale and a micrometric scale. We can mention a fourth system, the “Leveling system” , whose function is to level the analytical balance, so that the weighing result can be recorded correctly, thereby avoiding errors. It is made up of screw-type legs and a bubble.

Installation requirements

  1. Have an environment that does not present drafts, sudden temperature changes and that is free of dust. 2. Have a perfectly level table. A high inertia platform, isolated from the structures located in the vicinity, is ideal to reduce the effect of vibrations emitted by certain equipment such as centrifuges and refrigerators. It must have a sufficient area to install the balance and that the auxiliary equipment with which it interacts in the weighing processes. In the same way, the space required by the interconnection cables, electric current, connection to the connection system, to the printer, etc. must be foreseen. 3. Avoid installing equipment that produces high magnetic fields or vibrations such as centrifuges, electric motors, compressors and generators in the vicinity. Four. Avoid being under the direct influence of air conditioning systems – air currents – and sunlight. 5. Have an electrical outlet in good condition, equipped with an earth pole provided with switches, which complies with the electrical regulations in force in the country or in the laboratory.

Maintenance

The balance is characterized by being a high precision instrument. For this reason, the maintenance routines by the operator are minimal and are limited to the following:

  1. Clean the weighing pan so that it is free of dust or dirt. Cleaning is done with a clean piece of cloth that may be moistened with distilled water. If a stain needs to be removed, a mild detergent can be applied. A soft hair brush can also be used to remove particles or dust that may have been deposited on the weighing pan. 2. Clean externally and internally the weighing chamber. Verify that the glasses are free of dust. 3. Verify that the adjustment mechanisms on the front door of the weighing chamber are working properly.

Very important: Never lubricate a balance unless the manufacturer expressly indicates it. Any substance that interferes with the balance mechanisms delay its response or permanently alternate the measurement. In general, the manufacturer or the representative in specialized installations performs the maintenance of the scales, following procedures that vary depending on the type and model of the balance.

Handling of the Analytical Balance a) The weighing chamber is opened and cleaned internally with a camel hair brush. Once finished, the chamber doors are closed. b) Turn on the balance by pressing the control bar. The display rectangle is kept on for several seconds, and then set to 0.0000 g. (An automatic test is carried out). c) Open the glass doors of the weighing chamber and place the filter paper to weigh small masses on the weighing pan. It is also recommended to use a watch glass. The TLC shall wear gloves when handling materials. d) Close the glass doors by sliding them gently. Wait for the benchmark being the stability guide and indicates that the weight is stable. Tare the sample container. e) Open the weighing chamber again and carefully add the substance to be weighed to the desired mass, taking care to place the sample in the center of the container avoiding to throw the reagent out of the container area. f) Before registering the mass, close the glass doors and wait until the stability detector appears constant. g) Once the desired weight has been obtained. The weighing chamber must be left clean.

Do not take dishes from the pan with your bare hands as your fingerprints add mass. Use crucible tongs to prevent unwanted dough. It is recommended that the clamps be degreased with an alcohol swab. Do not lean on the bench while weighing. And if necessary write down the mass of your container, if you will require it for later.